Oxymitter 4000, Alarms eeprom corrupt, Instruction manual – Emerson Process Management OXYMITTER 4000 IM-106-340 User Manual

Page 110

Advertising
background image

8-18

Instruction Manual

IM-106-340, Rev. 4.0

May 2006

Oxymitter 4000

Figure 8-14. Fault 12, EEprom Corrupt

Fault 12, EEprom Corrupt

Figure 8-14 shows the electronic assembly for
an Oxymitter 4000 with a membrane keypad
(upper view) and an Oxymitter 4000 with an LOI
(lower view).

Membrane Keypad

When Fault 12 is detected, the O2 CELL LED
flashes four times, pauses for three seconds,
and repeats.

1. This alarm can occur if the EEprom is

changed for a later version. At power up,
the EEprom is not updated.

2. To correct this problem, power down and

then restore power. The alarm should clear.

3. If the alarm occurs while the unit is running,

there is a hardware problem on the
microprocessor board.

4. If cycling the power does not clear the

alarm, see "Electronic Assembly
Replacement" in
Section 9: Maintenance
and Service.

LOI

When Fault 12 is detected, the LOI displays the
"EEprom Corrupt" message. Refer to the
comments and procedures in paragraphs 1
through 4 above.

37260030

LOI

KEYPAD

DIAGNOSTIC

ALARMS

TEST

POINTS

HEATER T/C

HEATER

O2 CELL

CALIBRATION

CALIBRATION RECOMMENDED

O2 CELL mV +

O2 CELL mv -

HEATER T/C +

HEATER T/C -

INC

INC

DEC

DEC

HIGH

GAS

LOW

GAS

CAL

TEST GAS +

PROCESS -

% O2

SW2

TP1

J1

TP2

TP3

RED

YEL

GRN

ORG

TP4

TP5

TP6

ON

Alarms

EEprom Corrupt

Advertising