Figure 5-2), Oxymitter 4000, Instruction manual – Emerson Process Management OXYMITTER 4000 IM-106-340 User Manual

Page 71

Advertising
background image

Instruction Manual

IM-106-340, Rev. 4.0
May 2006

5-3

Oxymitter 4000

Figure 5-2. Calibration Keys

Keys

INC and DEC. The INC and DEC keys are used to set the values of the
calibration gases. Attach a multimeter across TP5 and TP6. The calibration
and process gases can now be monitored. Pressing the INC or DEC once will
cause the output to switch from the process gas to the calibration gas.
Pressing INC or DEC a second time will increase or decrease the calibration
gas parameter. If the keys have been inactive for one minute, the output
reverts to the process gas. When a calibration has been initiated, the value at
TP5 and TP6 is the % O

2

seen by the cell.

Oxygen levels, as seen on the multimeter, are:

8.0% O

2

= 8.0 volts DC

0.4% O

2

= 0.4 volts DC

CAL

The CAL key can:

• Initiate a calibration.
• Sequence through calibration.
• Abort the calibration.

NOTE
Refer Section 9: Maintenance and Service, for calibration instructions.

Model 751 Remote Powered Loop LCD Display (Optional)

Refer to Remote Powered Loop LCD manual for calibration and operation.

DIAGNOSTIC

ALARMS

TEST

POINTS

HEATER T/C

HEATER

02 CELL

CALIBRATION

CALIBRATION RECOMMENDED

02 CELL mV +
02 CELL mv -
HEATER T/C +
HEATER T/C -

INC

INC

DEC

DEC

HIGH

GAS

LOW

GAS

CAL

TEST GAS +

PROCESS -

% 02

Membrane

Keys

Membrane

Keys

Membrane

Key

Diagnostic

LEDs

22220023