Oxymitter 4000, Instruction manual – Emerson Process Management OXYMITTER 4000 IM-106-340 User Manual

Page 60

Advertising
background image

Oxymitter 4000

3-4

Instruction Manual

IM-106-340, Rev. 4.0

May 2006

Figure 3-2. Defaults - Oxymitter

4000 with Membrane Keypad

Read O

2

Concentration

Once the cell is up to operating temperature, the O

2

percentage can be read:

1. Access TP5 and TP6 next to the membrane keypad. Attach a

multimeter across TP5 and TP6. The calibration and process gases can
now be monitored. Pressing the INC or DEC once will cause the output
to switch from the process gas to the calibration gas. Pressing INC or
DEC a second time will increase or decrease the calibration gas
parameter. If the keys have been inactive for one minute, the output
reverts to the process gas. If the keys have been inactive for one
minute, the output reverts to the process gas.

37260010

DIAGNOSTIC

ALARMS

TEST

POINTS

HEATER T/C

HEATER

O2 CELL

CALIBRATION

CALIBRATION RECOMMENDED

O2 CELL mV +

O2 CELL mV -

HEATER T/C +

HEATER T/C -

INC

INC

DEC

DEC

HIGH

GAS

LOW

GAS

CAL

TEST GAS +

PROCESS -

% O2

SW2

SW1

TP1

J1

TP2

TP3

RED

YEL

GRN

ORG

TP4

TP5

TP6

ON

4-20 mA

is internally

powered (Default)

4-20 mA requires

external power

HART

0 to 10% O

2

3.5 mA

220 V

115 V

0 to 25% O

2

Local

21.6 mA

Default

position

(Ex-factory)

3.5 mA/21.6 mA:

0 to 25% O :

2

0 to 10% O /

2

Local:

HART: O Range set by HART/AMS

(From 0 to 40% O )
O Range set by Pos 2

O Range
When alarm exists, or on power-
up, output current goes to this
value

2

2

2

2

1

2
3
4

ON

OFF

Note:

The 115 V option at switch
SW2 position 4 is active only
when the heater voltage
option is set to manual in the
software
(auto tune = no).