Ent kit (figure 9-10), Oxymitter 4000, Instruction manual – Emerson Process Management OXYMITTER 4000 IM-106-340 User Manual

Page 133

Advertising
background image

Instruction Manual

IM-106-340, Rev. 4.0
May 2006

9-17

Oxymitter 4000

Figure 9-10. Cell Replacement Kitl

1. Follow the instructions in "Removal and Replacement of Probe" to

remove the Oxymitter 4000 from the stack or duct.

NOTE
To determine if the diffuser needs to be replaced, refer to "Calibration with
Keypad".

2. If the probe uses the standard diffusion element, use a spanner wrench

to remove the diffuser assembly.

35830009

Wire and
Pad Assembly

Anti-Seize
Compound

22 Ga.
Wire

Closed end
Connector

Set Screws

Teflon
Tubing

Socket Head
Cap Screws

Cell and
Flange
Assembly

Calibration Gas

Passage

ANSI

Gasket

Probe Tube
(Not Included in Kit)

Corrugated

Seal

Use heat-resistant gloves and clothing when removing the probe. Do not attempt to work on
these components until they have cooled to room temperature. Probe components can be
as hot as 572°F (300°C). This can cause severe burns.

Disconnect and lock out power before working on any electrical components. There is
voltage of up to 115 VAC.

Do not remove the cell unless certain it needs to be replaced. Removal may damage the
cell and platinum pad. Go through the complete troubleshooting procedure to make sure the
cell needs to be replaced before removing it.

Advertising