Oxymitter 4000

3-2

Instruction Manual

IM-106-340, Rev. 4.0

May 2006

Figure 3-1. Electronics Housing

Terminals and Membrane

Keypad

Oxymitter 4000
Configuration

Located on the microprocessor board, the top board, are two switches that
configure outputs for the Oxymitter 4000 (Figure 3-2). SW1 determines if the
4-20 mA signal is internally or externally powered. SW2 determines:

1. Oxymitter 4000 status, HART or LOCAL.
2. Oxygen range, 0 to 10% O

2

or 0 to 25% O

2

. (0 to 40% O

2

is also

configurable only through HART/AMS.)

3. The 4-20 mA signal, at fault or power up, 3.5 mA or 21.6 mA.

SW1 Setting

The two settings are internally or externally powering the 4-20 mA signal. The
factory setting is for the 4-20 mA signal to be internally powered.

37260009

AC

L1

AC

N

+

+

-

-

4-20

500 VA

SERIAL NO.
TAG NO.

OXYMITTER 4000

WATTS:

VOLTS:

FUSE:

LINE

OUTPUT:

Rosemount Analytical Inc.
Orrville, OH 44667-0901

85-264 VAC 48-62 Hz

TM

800-433-6076

4-20 mA

R

5 Amps

TM

HART

SMART FAMILY

DIAGNOSTIC

ALARMS

TEST

POINTS

HEATER T/C

HEATER
02 CELL

CALIBRATION

CALIBRATION RECOMMENDED

02 CELL mV +

02 CELL mv -

HEATER T/C +

HEATER T/C -

INC

INC

DEC

DEC

HIGH

GAS

LOW
GAS

CAL

TEST GAS +

PROCESS -

% 02

SW2

TP1

J1

TP2

TP3

RED

YEL

G

R

N

O

RG

TP4

TP5

TP6

O

N

4-20 mA
Signal

Logic I/O

Ground Lugs

Terminal
Block

Oxymitter 4000
Electronics
Housing

Remove power from the Oxymitter 4000 before changing defaults. If defaults are changed
under power, damage to the electronics package may occur.