Oxymitter 4000 setup at the loi, Oxymitter 4000 setup at the loi -6, Oxymitter 4000 – Emerson Process Management OXYMITTER 4000 IM-106-340 User Manual

Page 78: Instruction manual

Advertising
background image

Oxymitter 4000

6-6

Instruction Manual

IM-106-340, Rev. 4.0

May 2006

Figure 6-4. Menu Tree for Local

Operator Interface (LOI) on the

Oxymitter 4000 (Sheet 2 of 2)

OXYMITTER 4000 SETUP
AT THE LOI

In setting up the Oxymitter 4000 from the LOI, it is best to start at the
SYSTEM/Calibration Setup menu, Figure 6-4.

SYSTEM/Calibration Setup

O2 Gas #1 - Enter the high or low cal gas value (the order is not important).

O2 Gas #2 - Enter the second cal gas value.

NOTE
Refer to Section 9: Maintenance and Service, for calibration instructions.

SYSTEM

O2 Gas 1
O2 Gas 2
O2-Reset Vals
O2
O2 Cal Intervl
O2-
Gas Time
Purge Time

_____%
_____%

____H

___Sec.

Yes/No
Yes/No

____H

___Sec.

Yes/No

Out Tracks

Next Cal

Auto Calib?

O2 Type
O2 Range
O2 Alarm Leve

_______

______%

l

_____mA

Do O2 Trim

Calib Setup

Input/Output

Parameters

Status

Software

37260018

(CONTINUED FROM

SHEET 1)

Analog

Digital

NOTE

In column four of this menu, the selections in

are user configurable.

text selections are

procedures; related instructions are displayed on the LOI. All other parameters are display only.

Italics

Bold

Version

xxx

Checksum

xxx

Build Number

xxx

Build Date

xxxxxx

Test Code

xx

SW Err File

xx

SW Err Line

xx

Alarms

__________

PID Parameters

115/220

Yes/No

(Cal. required after reset)

Reset Device?

O2 Slope
O2 Constant
O2 T90 Time
Auto Tune?
Lockout Time
Revert Time
Luminance

____ mV/D

____ mV

0:00

Yes/No

0:00
0:00

______

Logic IO Mode
Low O2 Alarm

Input State
Force Output

See Table 4-1

See “Abort

Calibration” in Section 9.

Advertising