Instruction Manual

IM-106-340, Rev. 4.0
May 2006

11-3

Oxymitter 4000

IMPS 4000 INTELLIGENT
MULTIPROBE TEST GAS
SEQUENCER

Figure 11-3. IMPS 4000

The IMPS 4000 Intelligent Multiprobe Test Gas Sequencer is housed within
an IP56 (NEMA 4X) enclosure and has the intelligence to provide calibration
gas sequencing of up to four Oxymitter 4000 units to accommodate automatic
and semi-automatic calibration routines.

This sequencer works in conjunction with the Oxymitter 4000 CALIBRATION
RECOMMENDED feature, eliminating out-of-calibration occurrences and the
need to send a technician to the installation site. In addition, the SPS 4001B
provides a remote contact input to initiate a calibration from a remote location
and relay outputs to alert when a calibration is in progress, an Oxymitter 4000
is out of calibration, calibration gases are on, and calibration gas pressure is
low.

For more information, call Emerson Process Management at 1-800-433-6076.