R "cell replace- ment" in, Cell, Replacement" in – Emerson Process Management OXYMITTER 4000 IM-106-340 User Manual

Page 132: Cell replacement" in, Figure 9-9. make sur, S, figure 9-9. m, Oxymitter 4000, Instruction manual

Advertising
background image

Oxymitter 4000

9-16

Instruction Manual

IM-106-340, Rev. 4.0

May 2006

Figure 9-9. Probe to Probe Head

Assembly - Remote Electronics

Only

Cell Replacement

This paragraph covers oxygen sensing cell replacement. Do not attempt to
replace the cell until all other possibilities for poor performance have been
considered. If cell replacement is needed, order the cell replacement kit
(Table 10-1). Refer to Figure 9-3 or Figure 9-4 to view the component parts of
the Oxymitter 4000.

The cell replacement kit (Figure 9-10) contains a cell and flange assembly,
corrugated seal, setscrews, socket head cap screws, and anti-seize
compound. The items are carefully packaged to preserve precise surface
finishes. Do not remove items from the packaging until they are ready to be
used. Spanner wrenches and hex wrenches needed for this procedure are
part of an available special tools kit (Table 10-1 and Figure 10-2).

Probe Signal

Wire Connector

Probe Heater

Wire Connector

4 Wire

2 Wire

Housing

O-Ring

Probe

37260058

Adapter

Board