Calibration passes, but still reads incorrectly, Oxymitter 4000, Instruction manual – Emerson Process Management OXYMITTER 4000 IM-106-340 User Manual

Page 114

Advertising
background image

Oxymitter 4000

8-22

Instruction Manual

IM-106-340, Rev. 4.0

May 2006

CALIBRATION PASSES,
BUT STILL READS
INCORRECTLY

There are a few fault conditions where no alarm indication is present and the
probe passes calibration, but the O

2

reading may still be incorrect:

Probe passes calibration, but still appears to read high

There may be a leak that is permitting ambient air to mix with the process
gases. Since many combustion processes are slightly negative in pressure,
ambient air can be sucked into the cell area, biasing the O

2

reading upward.

1. Make sure that the calibration gas line is capped tightly between

calibrations. If autocal is used, make sure the check valve is seating
properly.

2. If an abrasive shield is installed to protect the entire probe from

particulate erosion, a leak in the probe flange gasket can allow ambient
air to migrate down the annular space between the probe and shield,
and then into the cell. Always install a new probe flange gasket when
reinstalling a probe.

There may be a leak inside the probe itself, permitting the reference air
(20.95% O

2

) to mix with the process gases at the cell. To confirm this leak

condition, instrument air will need to be connected for reference. Pressurize
the inside (reference side) of the probe by plugging the reference air exhaust
port with your finger for 1 minute. The O

2

reading should decrease slightly. If

the O

2

reading increases during this test, there is a leak inside the probe.

1. Acid condensation inside the probe can degrade the red silicon tube

(28, Figure 9-3) that carries the cal gas to the cell. Remove housing (21)
to inspect this hose. (See Section 9: Maintenance and Service). Black
vitan material is optionally available with greater chemical resistance.

Figure 8-18. Probe Leakage

Paths

2. The sensing cell is bolted to the end of the probe, and uses a corrugated

metallic seal (5, Figure 9-3) to separate the process gases from the
ambient reference air. This seal can be used only one time, so always
replace this seal when a cell is removed or replaced. Always apply
anti-seize compound on both sides of the corrugations.

BOTTOM VIEW

PROBE HEAD

Corrugated

Seal

Reference Air

Exhaust Port

Probe

Flange

Gasket

Advertising