Logic i/o, Logic i/o -4, Rmation, refer to "logic i/o – Emerson Process Management OXYMITTER 4000 IM-106-340 User Manual

Page 66: Oxymitter 4000, Instruction manual

Advertising
background image

Oxymitter 4000

4-4

Instruction Manual

IM-106-340, Rev. 4.0

May 2006

Figure 4-2. Defaults - Oxymitter

4000 with LOI

LOGIC I/O

This two-terminal logic contact can be configured either as a solid-state
relay-activated alarm or as a bi-directional calibration handshake signal to an
IMPS 4000 or SPS 4001B. The configuration of this signal depends on the
setting of the LOGIC I/O PIN MODE via HART/AMS or LOI. The ten different
modes available are explained in Table 4-1.

Alarm

When configured as an alarm, this signal alerts you to an out-of-spec
condition. The output is +5 Vdc in series with a 340 ohm resistor.

For optimum performance, Rosemount Analytical recommends connecting
the output to a Potter & Brumfield 3.2 mA DC relay (P/N R10S-E1Y1-J1.0K).

37260012

SW2

SW1

1

2
3
4

TP1

J1

TP2

TP3

RED

YEL

GRN

ORG

TP4

TP5

TP6

4-20 mA

is internally

powered
(Default)

4-20 mA requires

external power

HART

0 to 10% O

2

3.5 mA

220 V

115 V

0 to 25% O

2

Local

21.6 mA

Default

position

(Ex-factory)

3.5 mA/21.6 mA

0 to 25% O :

2

0 to 10% O /

2

Local:

HART: O Range set by HART/AMS

(From 0 to 40% O )
O Range set by Pos 2

O Range
When alarm exists, or
on power up, current
output goes to this value

2

2

2

2

ON

OFF

Note:

The 115 V option
at switch SW2
position 4 is active
only when the
heater voltage
option is set to
manual in the
software.