Table 8-1. Diagnostic/Unit

Alarm Fault Definitions -

Membrane Keypad Only

*Critical alarm conditions will render the O

2

measurement as unusable, and any of these events will cause the 4-20 mA signal to go to a

user-selectable limit of 3.5 mA or 21.6 mA (position 3 of SW2). Factory default value is 3.5 mA. Alarms which are not self-clearing
(Self-Clearing = NO) will require a reset. Perform the Reset Procedure in Section 3: Configuration of Oxymitter 4000 with Membrane
Keypad to continue oper
ation.

**The CALIBRATION RECOMMENDED alarm flashes the Calibration Recommended LED on the operator's keypad.

Table 8-2. Diagnostic/Unit

Alarm Fault Definitions - LOI

LED

Flashes

Status

4-20 mA Line

Fault

Recoverable

HEATER T/C

1
2
3
4

Open
Shorted
Reversed
A/D Comm Error

3.5 mA (factory default)*
3.5 mA (factory default)*
3.5 mA (factory default)*
3.5 mA (factory default)*

1
2
3
4

No
No
No
No

HEATER

1
2
3
4
5

Open
High High Temp
High Case Temp
Low Temp
High Temp

3.5 mA (factory default)*
3.5 mA (factory default)*
3.5 mA (factory default)*
3.5 mA (factory default)*
3.5 mA (factory default)*

5
6
7
8
9

No
No

Yes
Yes
Yes

O

2

CELL

1
3
4

High mV
Bad
EEprom Corrupt

3.5 mA (factory default)*
Track O

2

3.5 mA (factory default)*

10

11

12

Yes
Yes

No

CALIBRATION

1
2
3

**

Invalid Slope
Invalid Constant
Last Calibration Failed
Calibration Recommended

Track O

2

Track O

2

Track O

2

Track O

2

13
14
15

Yes
Yes
Yes
Yes

Message

Status

Fault Number

Self Clearing

O2 T/C OPEN

Heater T/C Open

1

No

O2 T/C SHORTED

Heater T/CShorted

2

No

O2 T/C REVERSED

Heater T/C Polarity Reversed

3

No

ADC ERROR

A/D Comm Error

4

No

O2 HEATER OPEN

O2 Heater Open

5

No

VERY HI O2 TEMP

Very High Process Temperature

6

No

BOARD TEMP HI

Electronics Overheated

7

Yes

O2 TEMP LOW

Low Process Temperature

8

Yes

O2 TEMP HI

High Process Temperature

9

Yes

O2 CELL OPEN

O2 Cell Open

10

Yes

O2 CELL BAD

O2 Cell Failed

11, 13, 14

Yes

EEPROM CORRUPT

EEprom Failed

12

No

CALIB FAILED

Last Calibration Failed

15

Yes

LINE FREQ ERROR

Incorrect Input Line Frequency Detected on Power Up

No