Figure 8-13. Fault 11, Bad Cell

Fault 11, Bad Cell

Figure 8-13 shows the electronic assembly for
an Oxymitter 4000 with a membrane keypad
(upper view) and an Oxymitter 4000 with an LOI
(lower view).

Membrane Keypad

When Fault 11 is detected, the O2 CELL flashes
three times, pauses for three seconds, and
repeats.

1. The bad cell alarm activates when the cell

exceeds the maximum resistance value.

2. The cell should be replaced. See "Cell

Replacement" in Section 9: Maintenance
and Service, for cell re
placement
instructions.

LOI

When Fault 11 is detected, the LOI displays the
"O2 Cell Bad" message. Refer to the comments
and procedures in paragraphs 1 and 2 above.