Oxymitter 4000

9-10

Instruction Manual

IM-106-340, Rev. 4.0

May 2006

Replace Entire Integral Electronics (with Housing)

NOTE
Only perform this procedure on units with integral electronics.

NOTE
Recalibration is required whenever electronic cards or sensing cell is
replaced.

1. Follow the instructions in "Removal and Replacement of Probe", to

remove the Oxymitter 4000 from the stack or duct.

2. Remove the right housing cover uncovering the electronic assembly

(Figure 9-5).

3. Depress and remove the J1 (cell and T/C) connector from the J1 socket.

Loosen the three captive mounting screws (16, Figure 9-3 or Figure 9-4)
on the microprocessor board (top board).

4. The J8 connector (heater leads) (Figure 9-6) can be accessed by

moving the J1 connector leads out of the slot on the microprocessor
board (17) and sliding the electronic assembly (12) partially out of the
housing (Figure 9-3 or Figure 9-4).

5. Squeeze the J8 connector on the sides and carefully remove. The

electronic assembly can now be completely removed from the housing.

6. Remove four screws (7, Figure 9-3) from the probe finned housing. The

probe and the electronic housing can now be separated.

7. When reinstalling or replacing the electronic housing, make sure that

O-ring (10) is in good condition. Place the J1 and J8 connectors in the
hole on the flat side of the electronic housing.

8. Hold the J1 and J8 connectors out and to the probe side of the

electronic housing. Make sure that the conduit port of the electronic
housing is on the same side as the CAL and REF gas ports. Replace
the four screws and tighten.

9. Reconnect the J8 connector to the power supply board. Make sure the

connector is secure.

10. Holding the J1 connector leads, slide the electronic assembly the rest of

the way into the housing. Align the electronic assembly so that it fits
flush on the pins. To ensure that it is flush, gently try to rotate the
electronics. If the electronics rotates, repeat the alignment.

11. Reconnect the J1 connector to the microprocessor board. Ensure the

connector is secure and tighten the three captive screws on the
microprocessor board (top board).

12. Replace the housing cover and ensure it is tight.
13. Follow the instructions in "Removal and Replacement of Probe" to

install the Oxymitter 4000 into the stack or duct.