Oxymitter 4000, Alarms o2 cell bad, Instruction manual – Emerson Process Management OXYMITTER 4000 IM-106-340 User Manual

Page 112

Advertising
background image

8-20

Instruction Manual

IM-106-340, Rev. 4.0

May 2006

Oxymitter 4000

Figure 8-16. Fault 14, Invalid Constant

Fault 14, Invalid Constant

Figure 8-16 shows the electronic assembly for
an Oxymitter 4000 with a membrane keypad
(upper view) and an Oxymitter 4000 with an LOI
(lower view).

Membrane Keypad

When Fault 14 is detected, the CALIBRATION
LED flashes twice, pauses for three seconds,
and repeats.

1. After a calibration has been performed, the

electronics calculates a cell constant value.

2. If the cell constant value is outside of the

range, -4 mV to 10 mV, the alarm will
activate. See "Calibration with Keypad" in
Section 9: Maintenance and Service, and

verify the last calibration was performed
correctly.

3. Power down the Oxymitter 4000 and

remove it from the stack.

4. Replace the cell per "Cell Replacement" in

Section 9: Maintenance and Service.

LOI

When Fault 14 is detected, the LOI displays the
"O2 Cell Bad" message. Refer to the comments
and procedures in paragraphs 1 through 4
above.

37260032

DIAGNOSTIC

ALARMS

TEST

POINTS

HEATER T/C

HEATER

O2 CELL

CALIBRATION

CALIBRATION RECOMMENDED

O2 CELL mV +

O2 CELL mv -

HEATER T/C +

HEATER T/C -

INC

INC

DEC

DEC

HIGH

GAS

LOW

GAS

CAL

TEST GAS +

PROCESS -

% O2

SW2

TP1

J1

TP2

TP3

RED

YEL

GRN

ORG

TP4

TP5

TP6

ON

LOI

KEYPAD

Alarms

O2 Cell Bad

Advertising