E figure 9-2, Oxymitter 4000, Instruction manual – Emerson Process Management OXYMITTER 4000 IM-106-340 User Manual

Page 120: Alarms manual calibration

Advertising
background image

Oxymitter 4000

9-4

Instruction Manual

IM-106-340, Rev. 4.0

May 2006

Figure 9-2. Inside Right Cover

2. Verify the calibration gas parameters are correct per "Calibration with

Keypad".

3. If performing a manual calibration with the CALIBRATION

RECOMMENDED LED off and the CAL LED off, start at step a.

4. If performing a manual calibration with the CALIBRATION

RECOMMENDED LED on and the CAL LED on, start at step b.
a. Push the CAL key. The CALIBRATION RECOMMENDED LED will

come on and the CAL LED will be on solid. If a multimeter is
attached across TP5 and TP6, the reading will display the
percentage of oxygen seen by the cell.

b. Push the CAL key. The CALIBRATION RECOMMENDED LED will

turn off and the CAL LED will flash continuously. The Oxymitter 4000
can be configured so that the 4-20 mA signal will hold the last value.
The default condition is for the output to track. A flashing LED
indicates that the Oxymitter 4000 is ready to accept the first
calibration gas.

c. Apply the first calibration gas. (Electronics will abort the calibration if

step 4 is not done within 30 minutes).

d. Push the CAL key; the CAL LED will be on solid. A timer is activated

to allow the calibration gas adequate time to flow (default time of five
minutes). When the timer times out, the Oxymitter 4000 has taken
the readings using the first calibration gas and the CAL LED will flash
continuously. The flashing indicates the Oxymitter 4000 is ready to
take readings using the second calibration gas.

SW2 DIP SWITCH

HART

LOCAL

0-10%

0-25%

4mA

20mA
NOT USED

NOT USED

LED

HEATER

O2 CELL

CALIBRATION

HEATER T/C

STATUS

OPEN

SHORTED

REVERSED

HIGH HIGH

TEMP

HIGH CASE

TEMP

LOW TEMP

HIGH TEMP

OPEN

BAD

EPROM

CORRUPT

INVALID SLOPE

INVALID

CONSTANT

FLASHES

1

2

3

4

1

2

OPEN

3

4

5

1

3

4

1

2

PUSH CAL

CAL LED ON

APPLY TG1

PUSH CAL

CAL LED ON SOLID

REMOVE TG1 & APPLY TG2

PUSH CAL

CAL LED ON SOLID

WAIT FOR FLASH

REMOVE TG2

PUSH CAL

1

2

3

4

5

6

7

ALARMS

MANUAL
CALIBRATION

CAL LED ON FOR
PURGE TIME

WAIT FOR FLASH

CAL LED OFF

PUSH CAL

CAL LED FLASH

2 FLASH-VALID CAL

3 FLASH-INVALID CAL

8

PLACE CONTROL LOOP
IN MANUAL
IF CAL LED ON
GO TO STEP 2

*
*

3

LAST CAL

FAILED

A/D COMM

ERROR