802BT/802R Video Test Generator User Guide

107

In the remote with lockout state (RWLS), the generator is under complete remote control

and front panel controls are disabled. The generator enters the RWLS state when the

controller issues the local lockout (LLO) message to the generator. Front panel access is

re-enabled when the controller issues the go to local (GTL) message to the generator.

In the local state (LOCS), the generator is under local control and all front panel controls

are enabled. Any remote messages received are stored for processing when the

generator enters the remote state again. The generator enters the remote state (REMS) if

the REN line is true and the generator is addressed to listen.

In the local with lockout state (LWLS), the generator is under local control, and all front

panel controls are enabled. Any remote messages received are stored for processing

when the generator enters the remote state again. The generator enters the remote with

lockout state (RWLS) if it is addressed to listen.