Avpg, Avpg analog video pedestal gate – Quantum Data 802BT User Manual

Page 251

Advertising
background image

Video Test Generator User Guide

241

AVPG

Analog Video Pedestal Gate

Class

Format parameter setting - Analog video signal

Description

Enables and disables the analog video set-up pedestal. The AVPG? query returns the

current setting of AVPG.

Command syntax

AVPG mode

mode

0 = OFF

1 = ON

Query syntax

AVPG?

Query response

mode

Other required

commands

Analog video must be enabled with the

AVST

command in order to output an analog video

signal. The pedestal level is set with the

AVPS

command. The

FMTU

command instructs

the generator to use the new setting. The

ALLU

command updates hardware to the new

setting and redraws the test image.

Example

AVPG 1

// Enable use of a black level pedestal

AVPS 7.5

// Set pedestal level to 7.5 IRE

FMTU

// Update hardware to current buffer contents