Testing analog computer (it) crts – Quantum Data 802BT User Manual

Page 41

Advertising
background image

802BT/802R Video Test Generator User Guide

31

Testing analog computer (IT) CRTs

This section describes how to test analog computer (IT) displays.

To test an analog computer CRT:

1. Use a standard VGA cable to connect the VGA connector on the generator with the

VGA connector on the display under test.

2. Calibrate the generator by holding down the

R, G

and

B

keys while starting the

generator. See “

Calibrating the generator

” on page 47.

You can also calibrate the generator by entering the following command:

SCAL

When calibrating the generator, if you load the analog outputs (as they will be loaded

when conducting testing) then the calibrations will be slightly more accurate. This is not

required however. Also, loading the unused outputs with 75 ohms will optimize signal

damping.

3. (Optional) Enable the Status Display mode and Analog Friendly mode (see

Special

operating modes

” on page 18). The Analog Friendly mode ensures that the generator

will output an analog signal. The Status Display mode enables you to view the current

output characteristics on the LCD.

4. Determine the formats to test (see

Selecting formats automatically

” on page 25).

5. (Optional) To view the EDID on the display, use the

Image

knob to select the EdidData

image and follow the directions on the display. You may wish to view the EDID data to

verify that the formats required for testing are specified in the EDID.

Alternatively, you can load the image with the following command:

IMGL

EdidData;

IMGU

6. Determine additional formats to test based on the resolution of the display. The VESA

formats are shown below:

Standard

Quantum Data format name

VGA

DMT06xx

SVGA

DMT08xx

XGA

DMT10xx

SXGA

DMT12xx

UXGA

DMT16xx

QXGA

DMT20xx

QSXBA

DMT25xx

Advertising