Testing hdmi sink devices for edid compliance – Quantum Data 802BT User Manual

Page 184

Advertising
background image

174

Chapter 10

Testing EDID

Testing HDMI sink devices for EDID compliance

The procedures for EDID and video testing of HDMI sink devices in accordance with the

HDMI Compliance Test Specification 1.0 are provided below.

Note: The EDID and video tests should be performed in sequence as a single test series.

Each test is dependent on successful completion of the previous test.

To set up the generator for HDMI compliance testing:

1. Hold down the

G

and

B

keys, and then power on the generator.

The following message appears on the front panel LCD.

2. Release the

G

and

B

keys.

The generator is placed in Digital Friendly mode.

3. Select and apply a valid HDMI format such as DMT0660 either by turning the Format

knob or entering the following command:

FMTL DMT0660; FMTU

4. Load the GenOps image to verify that the generator is configured to output HDMI (that

it is not in DVI-only mode).

The generator’s default configuration is to output HDMI. Note that the display status

mode will indicate that the generator is in HDMI mode with an “H” character. You can

also verify that the generator is putting out an HDMI signal by issuing the following

command:

HDMI?

The returned value should be 1 indicating HDMI output.