Vspp, Vspp vertical sync pulse polarity – Quantum Data 802BT User Manual

Page 576

Advertising
background image

566

Appendix A Command Reference

VSPP

Vertical Sync Pulse Polarity

Class

Format parameter setting - Synchronization

Description

Establishes the logic sense of the digital vertical sync outputs. Setting polarity to 1 causes

the leading edge of vertical sync to be a low-to-high transition. Setting polarity to 0 causes

the leading edge of vertical sync to be a high-to-low transition. The VSPP? query returns

the current polarity of VSPP.

Command syntax

VSPP polarity

polarity

0 = high-to-low transition

1 = low-to-high transition

Query syntax

VSPP?

Query response

0 or 1

Other required

commands

In order to use digital vertical sync, it must be gated on with the

VSPG

command and

digital separate H and V sync must be selected with the

SSST

command. The

FMTU

command instructs the generator to use the new setting. The

ALLU

command updates

hardware to the new setting and redraws the test image.

Example

VSPP 1 // Set active hi V sync in buffer
VSPG 1 // Enable V sync output in buffer
SSST 1 // Select H&V sync type in buffer
FMTU

// Update hardware to current buffer contents