Specifications, Approvals & Warranty
2.1.
Specifications
Table 2-1: Model 6400EH Basic Unit Specifications
Min/Max Range
(Physical Analog Output)
In 1 ppb increments from 10ppm to 5,000 ppm, dual
ranges or auto ranging
Measurement Units
ppm, mg/m3 (user selectable)
Zero Noise
1
0.05 ppm rms
Span Noise
1
< 0.5% of reading (above 50 ppm)
Lower Detectable Limit
2
0.1 ppm rms
Zero Drift (24 hours)
< 1 ppm
Zero Drift (7 days)
< 2 ppm
Span Drift (7 Days)
< 0.5% FS
Linearity
1 % of full scale
Precision
0.5% of reading
1
Temperature Coefficient
< 0.1% per
o
C
Voltage Coefficient
< 0.05% per V
Lag Time
1
5
sec
Rise/Fall Time
1
95% in < 30 sec
Sample Flow Rate
700 cm
3
/min. ±10%
Temperature Range
5-40
o
C
Humidity Range
0 - 95% RH, non-condensing
Dimensions H x W x D
7" x 17" x 23.5" (178 mm x 432 mm x 597 mm)
Weight, Analyzer
(Basic Configuration)
45 lbs (20.5 kg) w/internal pump
Weight, Pump Pack
16 lbs (7 kg)
AC Power Rating
100 V, 50/60 Hz (3.25A); 115 V, 60 Hz (3.0 A);
220 – 240 V, 50/60 Hz (2.5 A)
Environmental
Installation category (over-voltage category) II; Pollution
degree 2
Analog Outputs
Three (3) Outputs