Specifications, approvals & warranty, Specifications – Teledyne 6400EH - Sulfur Dioxide Analyzer User Manual

Page 17

Advertising
background image

UV Fluorescence SO2 Analyzer

Specifications/Warranty

Teledyne

Analytical

Instruments

5

Specifications, Approvals & Warranty

2.1.

Specifications

Table 2-1: Model 6400EH Basic Unit Specifications

Min/Max Range
(Physical Analog Output)

In 1 ppb increments from 10ppm to 5,000 ppm, dual
ranges or auto ranging

Measurement Units

ppm, mg/m3 (user selectable)

Zero Noise

1

0.05 ppm rms

Span Noise

1

< 0.5% of reading (above 50 ppm)

Lower Detectable Limit

2

0.1 ppm rms

Zero Drift (24 hours)

< 1 ppm

Zero Drift (7 days)

< 2 ppm

Span Drift (7 Days)

< 0.5% FS

Linearity

1 % of full scale

Precision

0.5% of reading

1

Temperature Coefficient

< 0.1% per

o

C

Voltage Coefficient

< 0.05% per V

Lag Time

1

5

sec

Rise/Fall Time

1

95% in < 30 sec

Sample Flow Rate

700 cm

3

/min. ±10%

Temperature Range

5-40

o

C

Humidity Range

0 - 95% RH, non-condensing

Dimensions H x W x D

7" x 17" x 23.5" (178 mm x 432 mm x 597 mm)

Weight, Analyzer
(Basic Configuration)

45 lbs (20.5 kg) w/internal pump

Weight, Pump Pack

16 lbs (7 kg)

AC Power Rating

100 V, 50/60 Hz (3.25A); 115 V, 60 Hz (3.0 A);
220 – 240 V, 50/60 Hz (2.5 A)

Environmental

Installation category (over-voltage category) II; Pollution
degree 2

Analog Outputs

Three (3) Outputs

Advertising