Teledyne 6400EH - Sulfur Dioxide Analyzer User Manual

Page 63

Advertising
background image

UV Fluorescence SO2 Analyzer

Troubleshooting

Teledyne

Analytical

Instruments

51

9. Set the HVPS fine adjustment switch to its maximum setting (F).
10. Turn the gain adjustment potentiometer clockwise to its

maximum setting.

11. Set the front panel display to show

STABIL

(see Section 6.2.1 of

the 6400E Manual)

12. Feed span gas into the analyzer.
13. Wait until the

STABIL

value is below 0.5 ppm,

Note: Use a span gas equal to 80% of the reporting range.

Example: for a reporting range of 200 ppm, use a span gas
of 160 ppm.

14. Scroll to the

OFFSET

function and record the value.

15. Scroll to the

NORM PMT

value.

Note: Do not overload the PMT by accidentally setting both

adjustment switches to their maximum setting.
This can cause permanent damage to the PMT.

16. Determine the target

NORM PMT

value according to the

following formulas.

If the reporting range is set for ≤ 500 ppm (the instrument
will be using the 500 ppm physical range):

Target

NORM PMT

= (8 x span gas concentration) +

OFFSET

If the reporting range is set for ≥ 2,001 ppb (the instrument
will be using the 5,000 ppm physical range):


Target

NORM PMT

= (0.8 x span gas concentration) +

OFFSET

EXAMPLE

: If the

OFFSET

is 33 mV, the Reporting Range is

1000 ppm, the span gas should be 800 PPM SO

2

and the

calculation would be:

Target

NORM PMT

= (0.8 x 800) + 33 mV

Target

NORM PMT

= 640 + 33 mV

Target

NORM PMT

= 673 mV

17. Set the

HVPS

coarse adjustment switch to the lowest setting that

will give you more than the target

NORM PMT

signal from Step

16.

Advertising