Maintenance

Teledyne

Analytical

Instruments

35

Instrument Maintenance

8.1 Maintenance Schedule

There is no Internal IZS offered for the 6400EH.

8.2 Predictive Diagnostics

Because the 6400EH’s internal pneumatics are monitored in a

different manner than those of the 6400E there are some differences in
how the instruments test functions are used as predictive diagnostics.
Table 8-1 of this addendum supersedes Table 9-2 of the 6400E Manual

Table 8-1: Predictive Uses for Test Functions

TEST

FUNCTION

IDAS

FUNCTION

CONDITION

BEHAVIOR

EXPECTED ACTUAL

INTERPRETATION

PRES

SMPPRS

Sample gas
pressure
u

pstream of the

critical flow
orifice.

Constant within
atmospheric
changes

Slowly
increasing

Slowly
decreasing

• Flow path is clogging up.

- Check critical flow
orifice & sintered filter.
- Replace particulate
filter

• Developing leak in

pneumatic system to
vacuum (developing
valve failure)

PRES

SMPPRS

Sample gas
pressure
u

pstream of the

critical flow
orifice.

Constant within
atmospheric
changes

Slowly
increasing

Slowly
decreasing

• Flow path is clogging up.

- Check critical flow
orifice & sintered filter.
- Replace particulate
filter

• Developing leak in

pneumatic system to
vacuum (developing
valve failure)

VAC

VACUUM

Gas pressure
downstream of
the critical flow
orifice (e.g.
inside reaction
cell.

Constant within
atmospheric
changes

Fluctuating

• Developing leak in

pneumatic system