Maintenance model 6400eh, Teledyne analytical instruments – Teledyne 6400EH - Sulfur Dioxide Analyzer User Manual

Page 48

Advertising
background image

Maintenance Model

6400EH

Teledyne Analytical Instruments

36

TEST

FUNCTION

IDAS

FUNCTION

CONDITION

BEHAVIOR

EXPECTED ACTUAL

INTERPRETATION

SAMP FL

SMPFLW

Standard

Operation

Stable

Slowly
Decreasing

• Flow path is clogging up.

- Check critical flow
orifice & sintered filter.
- Replace particulate
filter

DRK PMT

DRKPMT

PMT output
when UV Lamp
shutter closed

Constant within
±20 of check-
out value

Significantly
increasing

• PMT cooler failure

• Shutter Failure

SO2

Concentration

CONC1

Standard
configuration at
span

stable for
constant
concentration

Decreasing
over time

• Drift of instrument

response; UV Lamp
output is excessively
low.

Fluctuating

• Leak in gas flow path.

LAMP RATIO

LAMPR

Standard
Operation

Stable and near
100%

Fluctuating
or Slowly
increasing

• UV detector wearing out

• UV source Filter

developing pin holes

Slowly
deceasing

• UV detector wearing out

• Opaque oxides building

up on UV source Filter

• UV lamp aging