Maintenance model 6400eh, Teledyne analytical instruments – Teledyne 6400EH - Sulfur Dioxide Analyzer User Manual
Page 48

Maintenance Model
6400EH
Teledyne Analytical Instruments
36
TEST
FUNCTION
IDAS
FUNCTION
CONDITION
BEHAVIOR
EXPECTED ACTUAL
INTERPRETATION
SAMP FL
SMPFLW
Standard
Operation
Stable
Slowly
Decreasing
• Flow path is clogging up.
- Check critical flow
orifice & sintered filter.
- Replace particulate
filter
DRK PMT
DRKPMT
PMT output
when UV Lamp
shutter closed
Constant within
±20 of check-
out value
Significantly
increasing
• PMT cooler failure
• Shutter Failure
SO2
Concentration
CONC1
Standard
configuration at
span
stable for
constant
concentration
Decreasing
over time
• Drift of instrument
response; UV Lamp
output is excessively
low.
Fluctuating
• Leak in gas flow path.
LAMP RATIO
LAMPR
Standard
Operation
Stable and near
100%
Fluctuating
or Slowly
increasing
• UV detector wearing out
• UV source Filter
developing pin holes
Slowly
deceasing
• UV detector wearing out
• Opaque oxides building
up on UV source Filter
• UV lamp aging