Appendix C

Model 6400EH

Teledyne Analytical Instruments

86

Parameter

Displayed As

Observed

Value

Units

Nominal Range

Stray Light

STR. LGT

PPM

-50 to +100

Dark PMT

DRK PMT

MV

<200

Dark Lamp

DRK LMP

MV

-30 to 50

Slope

SLOPE

-

1.0 ± 0.3

Offset

OFFSET

MV

<200

High Voltage Power

Supply

HVPS

V

400 – 750*

Reaction Cell

Temperature

RCELL

TEMP

°C

50 ± 1

Box Temperature

BOX TEMP

°C

Ambient + (3-7)

PMT Temperature

PMT TEMP

°C

7 ± 2

Time of Day

TIME

HH:MM:SS


Test Settings

Test Value

Observed Value

Acceptable Value

ETEST PMT Reading

2000 ± 1000MV

OTEST PMT Reading

2000 ± 20 MV