Appendix c warranty questionnaire – Teledyne 6400EH - Sulfur Dioxide Analyzer User Manual
Page 97

UV Fluorescence SO2 Analyzer
Appendix C
Teledyne
Analytical
Instruments
85
Appendix C Warranty Questionnaire
Parameter
Displayed As
Observed
Value
Units
Nominal Range
Range
RANGE
PPM
UG/M3
1-5000 PPM Standard
Stability
STABIL
PPM
UG/M3
<.1 PPM with Zero Air
Vacuum
VACUUM
“Hg
4 – 10 “Hg
Sample Pressure
PRES
In-Hg-A
24 – 29
Sample Flow
SAMP FL
CC/MIN 700
±10%
PMT Signal
PMT
MV
0 ± 100 with Zero Air
Normalized PMT Signal NORM PMT
MV
0 ± 100 with Zero Air
UV Lamp
UV LAMP
MV
1000 – 4800
UV Lamp Ratio
LAMP
RATIO
%
35 – 120%
CUSTOMER: ______________________________ PHONE: ______________________
CONTACT NAME: _________________________ FAX NO. _____________________
SITE ADDRESS: ________________________________________________________
________________________________________________________
________________________________________________________
MODEL 6400EH SERIAL NO.: _______________ FIRMWARE REVISION: ________
1. ARE THERE ANY FAILURE MESSAGES? ________________________________
_______________________________________________________________________
_______________________________________________________________________
PLEASE COMPLETE THE FOLLOWING TABLE: (NOTE:
DEPENDING ON OPTIONS INSTALLED,
NOT ALL TEST PARAMETERS SHOWN BELOW WILL BE AVAILABLE IN YOUR
INSTRUMENT)