Appendix C

Teledyne

Analytical

Instruments

85

Appendix C Warranty Questionnaire

Parameter

Displayed As

Observed

Value

Units

Nominal Range

Range

RANGE

PPM

UG/M3

1-5000 PPM Standard

Stability

STABIL

PPM

UG/M3

<.1 PPM with Zero Air

Vacuum

VACUUM

“Hg

4 – 10 “Hg

Sample Pressure

PRES

In-Hg-A

24 – 29

Sample Flow

SAMP FL

CC/MIN 700

±10%

PMT Signal

PMT

MV

0 ± 100 with Zero Air

Normalized PMT Signal NORM PMT

MV

0 ± 100 with Zero Air

UV Lamp

UV LAMP

MV

1000 – 4800

UV Lamp Ratio

LAMP

RATIO

%

35 – 120%

CUSTOMER: ______________________________ PHONE: ______________________
CONTACT NAME: _________________________ FAX NO. _____________________
SITE ADDRESS: ________________________________________________________

________________________________________________________
________________________________________________________

MODEL 6400EH SERIAL NO.: _______________ FIRMWARE REVISION: ________
1. ARE THERE ANY FAILURE MESSAGES? ________________________________
_______________________________________________________________________
_______________________________________________________________________

PLEASE COMPLETE THE FOLLOWING TABLE: (NOTE:

DEPENDING ON OPTIONS INSTALLED,

NOT ALL TEST PARAMETERS SHOWN BELOW WILL BE AVAILABLE IN YOUR
INSTRUMENT)