Appendix A
Model 6400EH
Teledyne Analytical Instruments
70
NAME
MESSAGE TEXT
DESCRIPTION
WFRONTPANEL
FRONT PANEL WARN
Firmware is unable to communicate with the front panel.
WANALOGCAL
ANALOG CAL WARNING
The A/D or at least one D/A channel has not been
calibrated.
Table A-3: 6400EH Test Functions, Revision C.0
TEST FUNCTION
MESSAGE TEXT
DESCRIPTION
RANGE
RANGE=500.0 PPB
D/A range in single or auto-range modes.
RANGE1
RANGE1=500.0 PPB
D/A #1 range in independent range mode.
RANGE2
RANGE2=500.0 PPB
D/A #2 range in independent range mode.
STABILITY
STABIL=0.0 PPB
Concentration stability (standard deviation based on
setting of STABIL_FREQ and STABIL_SAMPLES).
VACUUM
VAC=9.1 IN-HG-A
Vacuum pressure.
SAMPPRESS
PRES=29.9 IN-HG-A
Sample pressure.
SAMPFLOW
SAMP FL=700 CC/M
Sample flow rate.
PMTDET
PMT=762.5 MV
Raw PMT reading.
NORMPMTDET
NORM PMT=742.9 MV
PMT reading normalized for temperature, pressure,
auto-zero offset, but not range.
UVDET
UV LAMP=3457.6 MV
UV lamp reading.
LAMPRATIO
LAMP RATIO=100.0 %
UV lamp ratio of current reading divided by calibrated
reading.
STRAYLIGHT
STR. LGT=0.1 PPB
Stray light offset.
DARKPMT
DRK PMT=19.6 MV
PMT dark offset.
DARKLAMP
DRK LMP=42.4 MV
UV lamp dark offset.
SLOPE
SLOPE=1.061
Slope for current range, computed during zero/span
calibration.
OFFSET
OFFSET=250.0 MV
Offset for current range, computed during zero/span
calibration.
HVPS
HVPS=650 VOLTS
High voltage power supply output.
RCELLDUTY
RCELL ON=0.00 SEC
Reaction cell temperature control duty cycle.
RCELLTEMP
RCELL TEMP=52.1 C
Reaction cell temperature.
BOXTEMP
BOX TEMP=35.5 C
Internal chassis temperature.
PMTTEMP
PMT TEMP=7.0 C
PMT temperature.
IZSDUTY
IZS ON=0.00 SEC
IZS temperature control duty cycle.