Appendix a model 6400eh, Teledyne analytical instruments – Teledyne 6400EH - Sulfur Dioxide Analyzer User Manual

Page 86

Advertising
background image

Appendix A

Model 6400EH

Teledyne Analytical Instruments

74

SIGNAL NAME

BIT OR CHANNEL

NUMBER

DESCRIPTION

1 = all pressures OK

ST_TEMP_ALARM 4

0 = any temperature alarm

1 = all temperatures OK

ST_HVPS_ALARM 5

0 = HVPS alarm

1 = HVPS OK

6–7 Spare

Front panel I2C keyboard, default I2C address 4E hex

MAINT_MODE 5

(input)

0 = maintenance mode

1 = normal mode

LANG2_SELECT 6

(input)

0 = select second language

1 = select first language (English)

SAMPLE_LED 8

(output)

0 = sample LED on

1 = off

CAL_LED 9

(output)

0 = cal. LED on

1 = off

FAULT_LED 10

(output)

0 = fault LED on

1 = off

AUDIBLEBEEPER 14

(output)

0 = beeper on (for diagnostic testing
only)

1 = off

Relay board digital output (PCF8575), default I2C address 44 hex

RELAY_WATCHDOG

0

Alternate between 0 and 1 at least every
5 seconds to keep relay board active

RCE LL_HEATER

1

0 = reaction cell heater on

1 = off

2–3 Spare

IZSHEATER 4

0 = IZS heater on

1 = off

5 Spare

CAL VALVE

6

0 = let cal. gas in

1 = let sample gas in

Advertising