Teledyne 6400EH - Sulfur Dioxide Analyzer User Manual

Page 42

Advertising
background image

Operating Instructions

Model 6400EH

Teledyne Analytical Instruments

30

Table 6-6: Control Input Signals

INPUT #

STATUS DEFINITION

ON CONDITION

A

REMOTE ZERO CAL

The analyzer is placed in Zero Calibration mode. The mode field of the
display will read ZERO CAL R.

B

REMOTE

HIGH SPAN CAL

If the instrument is set for DUAL or AUTO reporting rang mode,
activating this input causes the analyzer to enter high range span
calibration mode. The mode field of the display will read SPAN CAL R.

C

REMOTE

LO SPAN CAL

The analyzer is placed in low span calibration mode as part of
performing a low span (midpoint) calibration. The mode field of the
display will read LO CAL R.

D, E & F

SPARE

Digital Ground

The ground level from the analyzer’s internal DC power supplies
(same as chassis ground)

U

External Power input

Input pin for +5 VDC required to activate pins A – F.

+

5 VDC output

Internally generated 5V DC power. To activate inputs A – F, place a
jumper between this pin and the “U” pin. The maximum amperage
through this port is 300 mA (combined with the analog output supply, if
used).