Appendix a model 6400eh, Teledyne analytical instruments – Teledyne 6400EH - Sulfur Dioxide Analyzer User Manual

Page 74

Advertising
background image

Appendix A

Model 6400EH

Teledyne Analytical Instruments

62

Appendix A-2: Setup Variables For Serial I/O,
Revision C.0

Table A-1: 6400EH Setup Variables, Revision C.0

SETUP VARIABLE

NUMERIC

UNITS

DEFAULT

VALUE

VALUE RANGE

DESCRIPTION

DAS_HOLD_OFF

Minutes

15

0.5–20

Duration of DAS hold off period.

TPC_ENABLE —

ON

OFF,

ON

ON enables temperature and
pressure compensation; OFF
disables it.

RCELL_SET

ºC

50

30-70

Reaction cell temperature set point
and warning limits.

Warnings

45–55

DYN_ZERO —

OFF OFF,

ON

ON enables contact closure dynamic
zero; OFF disables it.

DYN_SPAN —

OFF OFF,

ON

ON enables contact closure dynamic
span; OFF disables it.

CONC_PRECISION —

1

AUTO, 0, 1, 2,

3, 4

Number of digits to display to the
right of the decimal point for
concentrations on the display.
Enclose value in double quotes (“)
when setting from the RS-232
interface.

CLOCK_ADJ

Sec./Day

0

-60–60

Time-of-day clock speed adjustment.

LANGUAGE_SELECT —

ENGL

ENGL, SECD,

EXTN

Selects the language to use for the
user interface. Enclose value in
double quotes (“) when setting from
the RS-232 interface.

MAINT_TIMEOUT

Hours 2 0.1–100

Time until automatically switching
out of software-controlled
maintenance mode.

CONV_TIME —

33

MS

33 MS, 66 MS,

133 MS 266 MS

533 MS, 1 SEC,

2 SEC

Conversion time for PMT and UV
detector channels. Enclose value in
double quotes (“) when setting from
the RS-232 interface.

DWELL TIME

_

Seconds 1

0.1–10

Dwell time before taking each
sample.

FILT_SIZE

Samples

30

1–480

Moving average filter size.

FILT_ASIZE

Samples 6

1–100

Moving average filter size in
adaptive mode.

FILT_DELTA

PPM 10 1–100

Absolute change to trigger adaptive
filter.

FILT_PCT

% 5 1–100

Percent change to trigger adaptive

Advertising