Appendix A
Model 6400EH
Teledyne Analytical Instruments
62
Appendix A-2: Setup Variables For Serial I/O,
Revision C.0
Table A-1: 6400EH Setup Variables, Revision C.0
SETUP VARIABLE
NUMERIC
UNITS
DEFAULT
VALUE
VALUE RANGE
DESCRIPTION
DAS_HOLD_OFF
Minutes
15
0.5–20
Duration of DAS hold off period.
TPC_ENABLE —
ON
OFF,
ON
ON enables temperature and
pressure compensation; OFF
disables it.
RCELL_SET
ºC
50
30-70
Reaction cell temperature set point
and warning limits.
Warnings
45–55
DYN_ZERO —
OFF OFF,
ON
ON enables contact closure dynamic
zero; OFF disables it.
DYN_SPAN —
OFF OFF,
ON
ON enables contact closure dynamic
span; OFF disables it.
CONC_PRECISION —
1
AUTO, 0, 1, 2,
3, 4
Number of digits to display to the
right of the decimal point for
concentrations on the display.
Enclose value in double quotes (“)
when setting from the RS-232
interface.
CLOCK_ADJ
Sec./Day
0
-60–60
Time-of-day clock speed adjustment.
LANGUAGE_SELECT —
ENGL
ENGL, SECD,
EXTN
Selects the language to use for the
user interface. Enclose value in
double quotes (“) when setting from
the RS-232 interface.
MAINT_TIMEOUT
Hours 2 0.1–100
Time until automatically switching
out of software-controlled
maintenance mode.
CONV_TIME —
33
MS
33 MS, 66 MS,
133 MS 266 MS
533 MS, 1 SEC,
2 SEC
Conversion time for PMT and UV
detector channels. Enclose value in
double quotes (“) when setting from
the RS-232 interface.
DWELL TIME
_
Seconds 1
0.1–10
Dwell time before taking each
sample.
FILT_SIZE
Samples
30
1–480
Moving average filter size.
FILT_ASIZE
Samples 6
1–100
Moving average filter size in
adaptive mode.
FILT_DELTA
PPM 10 1–100
Absolute change to trigger adaptive
filter.
FILT_PCT
% 5 1–100
Percent change to trigger adaptive