Calibration and Check

Teledyne

Analytical

Instruments

31

Calibration and Calibration Check Procedures

Calibration procedures for the 6400EH are the same as those for the

6400E. One exception to this statement is that the 6400EH has a special
valve option, Zero and Two Span Point Valve Option - OPT 52 (See
Section 5.1), that allows a mid-span point be checked.

7.1 Manual Calibration with the Zero and Two Span

Point Valve Option (OPT 52) installed

Note: It is only possible to calibrate to the high span gas. The low

span gas is only used for calibration checks.

Zero and Span calibrations using the Zero and two Span Valve

option are similar to that described in Section 7.2, except that:

 Zero air and both span gas is supplied to the analyzer through the

zero gas and span gas inlets rather than through the sample inlet.

 The zero and cal operations are initiated directly and

independently with dedicated keys (

CALZ

&

CALS

)

STEP ONE: Connect the sources of zero air and span gas to the
respective ports on the rear panel (see Figure 3-2 of this manual).

STEP TWO: Set the expected SO

2

high span gas value: