Ce mark compliance – Teledyne 6400EH - Sulfur Dioxide Analyzer User Manual
Page 18

Specifications/Warranty Model
6400EH
Teledyne Analytical Instruments
6
Analog Output Ranges
100 mV, 1 V, 5 V, 10 V, 2-20 or 4-20 mA isolated
current loop.
All Ranges with 5% Under/Over Range
Analog Output Resolution
1 part in 4096 of selected full-scale voltage
Status Outputs
8 Status outputs from opto-isolators
Control Inputs
6 Control Inputs, 3 defined, 3 spare
Serial I/O
One (1) RS-232; One (1) RS-485 (2 connecters in
parallel)
Baud Rate : 300 – 115200: Optional Ethernet Interface
Certifications
EN61326 (1997 w/A1: 98) Class A,
FCC Part 15 Subpart B Section 15.107 Class A,
ICES-003 Class A (ANSI C63.4 1992) & AS/NZS 3548
(w/A1 & A2; 97) Class A.
IEC 61010-1:90 + A1:92 + A2:95,
1 As defined by the USEPA.
2 Defined as twice the zero noise level by the USEPA.
2.2. CE Mark Compliance
Emissions Compliance
The TAI UV Fluorescence SO
2
Analyzer 6400EH was tested and
found to be fully compliant with:
EN61326 (1997 w/A1: 98) Class A, FCC Part 15 Subpart B Section
15.107 Class A, ICES-003 Class A (ANSI C63.4 1992) & AS/NZS
3548 (w/A1 & A2; 97) Class A.
Tested on 21 February 2003 - 08 March 2003 at CKC Laboratories,
Inc., Report Number CE03-021 A.
Safety Compliance
The TAI UV Fluorescence SO
2
Analyzer 6400EH was tested and
found to be fully compliant with:
I EC 61010-1:90 + A1:92 + A2:95,
Issued by CKC Laboratories on 4 April 2003, Report Number WO-
80146.