UV Fluorescence SO2 Analyzer

Troubleshooting

Teledyne

Analytical

Instruments

43

Troubleshooting & Repair

For the most part the information contained in Chapter 11 of the

6400E Manual is also applicable to the 6400EH. There are a few
exceptions however.

10.1.1 Fault Diagnosis with Warning Messages

Table 10-1: Warning Messages - Indicated Failures

WARNING

MESSAGE

FAULT CONDITION

POSSIBLE CAUSES

VACUUM

PRESS

WARN

Gas pressure inside the
reaction cell outside of
warning limits.

If sample pressure is > 10 in-Hg:
o

Pneumatic Leak

o

Bad Pump→Rebuild Pump

o

Failed pressure sensor/circuitry

10.1.2 Fault Diagnosis with Test Functions

Table 10-2: Test Functions - Possible Causes for Out-Of-Range Values

TEST

FUNCTION

NOMINAL
VALUE(S)

POSSIBLE CAUSE(S)

VAC

<

9.1 IN-HG-A

Incorrect sample gas pressure could be due to: pneumatic leak;
malfunctioning valve; malfunctioning pump; clogged flow orifices;
sample inlet overpressure; faulty pressure sensor

10.2 Subsystem Checkout

10.2.1 Pneumatic Sensor Assembly

The pneumatic sensor assembly of the 6400EH differs from that of

the 6400E in that there is no flow sensor. Instead the assembly includes
two pressure sensors located on either side of a critical flow orifice. The
6400EH software infers the gas flow rate by mathematically comparing
the two pressure readings.