Teledyne 6400EH - Sulfur Dioxide Analyzer User Manual

Page 52

Advertising
background image

Theory of Operation

Model 6400EH

Teledyne Analytical Instruments

40

9.2.2 Pneumatic Sensors

The 6400EH uses two pneumatic sensors to verify gas flow. These

sensors are located on a printed circuit assembly, called the pneumatic
pressure/flow sensor board. This PCA is attached to a manifold
containing the critical flow orifice that sets the instrument flow rate.

9.2.2.1

S

AMPLE

P

RESSURE

S

ENSOR

An absolute pressure transducer plumbed to the input of the

analyzer’s sample chamber is used to measure the pressure of the sample
gas before it passes through the critical flow orifice. This is used to
validate the critical flow condition (2:1 pressure ratio) through the
instrument’s critical flow orifice.

The actual sample gas pressure measurement is viewable through

the analyzer’s front panel display as the test function

PRES

.

9.2.2.2

V

ACUUM

P

RESSURE

S

ENSOR

An absolute pressure transducer measures the pressure on the

vacuum side of the critical flow orifice and is used to measure the
sample gas pressure in the reaction cell. If the vacuum pressure is not in
the correct range, a warning will be displayed by the software. Also, if
the temperature/pressure compensation (TPC) feature is turned on, the
output of this sensor is also used to supply pressure data for that
calculation.

The actual pressure of the gas downstream from the critical flow

orifice (including the gas inside the reaction cell) viewable through the
analyzer’s front panel display as the test function

VAC

.

9.2.2.3

S

AMPLE

F

LOW

C

ALCULATION

Unlike the 6400E, which uses a thermal-mass flow sensor to

directly measure the gas flow though the instrument, the 6400EH
calculates the gas as follows.

 The ratio of the two pressures is measured and used to validate

critical flow. If the ratio is not correct (< 2:1) the

SAMPLE

FLOW WARN

message is activated. Also, the value of the

SAMP FL

test function is set to

XXXX

.

If the pressure ratio between the two sensors is valid

(≥ 2:1),

the

instrument calculates the flow based on sample gas pressure level

(PRES)

and is viewable via the front panel as the

SAMP FL

test function.