Appendix A

Model 6400EH

Teledyne Analytical Instruments

66

SETUP VARIABLE

NUMERIC

UNITS

DEFAULT

VALUE

VALUE RANGE

DESCRIPTION

PRESSURE,

SAMPLE
PRESSURE,

SAMPLE FLOW,

RCELL TEMP,

CHASSIS
TEMP,

PMT TEMP,

HVPS
VOLTAGE

REMOTE_CAL_MODE

— LOW

LOW,

HIGH

Range to calibrate during contact-
closure and Hessen calibration.

Enclose value in double quotes (“)
when setting

PASS_ENABLE

— OFF

OFF,

ON

ON enables passwords; OFF
disables them.

STABIL_FREQ

Seconds 10

1–300

Stability measurement sampling
period.

STABIL_SAMPLES

Samples 25

2–40

Number of samples in concentration
stability reading.

RCELL CYCLE

_

Seconds 2

0.5–30

Reaction cell temperature control
cycle period.

RCELL_PROP

1/ºC

0.3

(prop. band

=

3 .3°C)

0-10

Reaction cell temperature PID
proportional coefficient.

RCELL INTEG

_

— 0.005 0–10

Reaction cell temperature PID
integral coefficient.

RCELL_DERIV

— 0.5 0–10

Reaction cell temperature PID
derivative coefficient.

HVPS_SET

Volts

550

0–2000

High voltage power supply warning
limits. Set point is not used.

Warnings:

400–700

DETECTOR_LIMIT

mV

1000

0–5000

UV lamp and PMT detector warning
limits. Set point is not used.

Warnings:

600-4995

SERIAL_NUMBER

— “00000000"

Any character in
the allowed
character set.
Up to 100
characters long.

Unique serial number for instrument.

Enclose value in double quotes (“)
when setting from the RS-232
interface.