Appendix a model 6400eh, Teledyne analytical instruments – Teledyne 6400EH - Sulfur Dioxide Analyzer User Manual
Page 78

Appendix A
Model 6400EH
Teledyne Analytical Instruments
66
SETUP VARIABLE
NUMERIC
UNITS
DEFAULT
VALUE
VALUE RANGE
DESCRIPTION
PRESSURE,
SAMPLE
PRESSURE,
SAMPLE FLOW,
RCELL TEMP,
CHASSIS
TEMP,
PMT TEMP,
HVPS
VOLTAGE
REMOTE_CAL_MODE
— LOW
LOW,
HIGH
Range to calibrate during contact-
closure and Hessen calibration.
Enclose value in double quotes (“)
when setting
PASS_ENABLE
— OFF
OFF,
ON
ON enables passwords; OFF
disables them.
STABIL_FREQ
Seconds 10
1–300
Stability measurement sampling
period.
STABIL_SAMPLES
Samples 25
2–40
Number of samples in concentration
stability reading.
RCELL CYCLE
_
Seconds 2
0.5–30
Reaction cell temperature control
cycle period.
RCELL_PROP
1/ºC
0.3
(prop. band
=
3 .3°C)
0-10
Reaction cell temperature PID
proportional coefficient.
RCELL INTEG
_
— 0.005 0–10
Reaction cell temperature PID
integral coefficient.
RCELL_DERIV
— 0.5 0–10
Reaction cell temperature PID
derivative coefficient.
HVPS_SET
Volts
550
0–2000
High voltage power supply warning
limits. Set point is not used.
Warnings:
400–700
DETECTOR_LIMIT
mV
1000
0–5000
UV lamp and PMT detector warning
limits. Set point is not used.
Warnings:
600-4995
SERIAL_NUMBER
— “00000000"
Any character in
the allowed
character set.
Up to 100
characters long.
Unique serial number for instrument.
Enclose value in double quotes (“)
when setting from the RS-232
interface.