Troubleshooting Model

6400EH

Teledyne Analytical Instruments

52

18. The coarse adjustment typically increments the

NORM PMT

signal in 100-300 mV steps.

19. Adjust the

HVPS

fine adjustment such that the

NORM PMT

value

is at or just above the target

NORM PMT

signal from Step 16.

20. Continue adjusting the both the coarse and fine switches until

NORM PMT

is as close to (but not below) the target

NORM PMT

value from Step 16.

21. Adjust gain adjustment potentiometer until the

NORM PMT

value

is ±10 mV of the target level from Step 16.

22. Perform span calibration (see Chapter 7 of the 6400E Manual)
23. Scroll to the

SLOPE

function and record the value.

24. If the value of the

SLOPE

is between 0.900 and 1.100 the PMT

Hardware calibration is complete.

25. If the value of the

SLOPE

is less than 0.900 or greater than 1.100:

a. Multiply the

slope

value from step 22 by the

norm PMT

value from step 19.

b. Repeat steps 17 through 24 using this new value for

NORM

PMT

.