4 technical assistance – Teledyne 6400EH - Sulfur Dioxide Analyzer User Manual
Page 64

Troubleshooting Model
6400EH
Teledyne Analytical Instruments
52
18. The coarse adjustment typically increments the
NORM PMT
signal in 100-300 mV steps.
19. Adjust the
HVPS
fine adjustment such that the
NORM PMT
value
is at or just above the target
NORM PMT
signal from Step 16.
20. Continue adjusting the both the coarse and fine switches until
NORM PMT
is as close to (but not below) the target
NORM PMT
value from Step 16.
21. Adjust gain adjustment potentiometer until the
NORM PMT
value
is ±10 mV of the target level from Step 16.
22. Perform span calibration (see Chapter 7 of the 6400E Manual)
23. Scroll to the
SLOPE
function and record the value.
24. If the value of the
SLOPE
is between 0.900 and 1.100 the PMT
Hardware calibration is complete.
25. If the value of the
SLOPE
is less than 0.900 or greater than 1.100:
a. Multiply the
slope
value from step 22 by the
norm PMT
value from step 19.
b. Repeat steps 17 through 24 using this new value for
NORM
PMT
.