Teledyne 6400EH - Sulfur Dioxide Analyzer User Manual
Page 75

UV Fluorescence SO2 Analyzer
Appendix A
Teledyne
Analytical
Instruments
63
SETUP VARIABLE
NUMERIC
UNITS
DEFAULT
VALUE
VALUE RANGE
DESCRIPTION
filter.
FILT_DELAY
Seconds 180
0–300
Delay before leaving adaptive filter
mode.
FILT_ADAPT —
ON OFF,
ON
ON enables adaptive filter; OFF
disables it.
DIL_FACTOR
—
1 0.1–1000
Dilution factor if dilution enabled with
FACTORY_OPT variable.
USER_UNITS —
PPM PPM,
UGM
Concentration units for user
interface. Enclose value in double
quotes (“) when setting from the RS-
232 interface.
LAMP CAL
_
mV
3500
1000–5000
Last calibrated UV lamp reading.
LAMP GAIN
—
0.9 0.5–1.5
UV lamp compensation attenuation
factor.
TEMPCO_GAIN —
0 0–2
Temperature coefficient attenuation
factor for pressure readings.
SLOPE CONST
—
6.25 0.1–10
Constant to make visible slope close
to 1.
DARK_ENABLE —
ON OFF,
ON
ON enables PMT/UV dark
calibration; OFF disables it.
DARK_FREQ
Minutes
30,
0.1–1440
Dark calibration period.
DARK LAMP OFF
Seconds
1
0.01–10
Dark calibration lamp off period.
DARK_PRE_DWELL
Seconds 10
1–60
Dwell time after closing dark shutter
or turning off lamp or selecting
preamp range.
DARK POST DWELL
Seconds 30
1–180
Dwell time after opening dark shutter
or turning on lamp.
DARK_SAMPLES
Samples
5
1–10
Number of dark samples to average.
DARK_FSIZE
Samples 2
1–100
Dark offset moving average filter
size.
DARK_LIMIT
mV
400
0–1000
Maximum dark offset allowed.
SO2 SPAN1
_
Conc 4000 0.1–50000
Target SO2 concentration during
span calibration of range 1.
SO2_SLOPE1
PPM/mV
1
0.25–4
SO2 slope for range 1.
SO2_OFFSET1
mV
0
-1500–1500
SO2 offset for range 1.
SO2 SPAN2
Conc 4000 0.1–50000
Target SO2 concentration during
span calibration of range 2.
SO2_SLOPE2
PPM/mV
1
0.25–4
SO2 slope for range 2.
SO2_OFFSET2
mV
0
-1500–1500
SO2 offset for range 2.
RANGE MODE
—
SNGL SNGL,
DUAL,
Range control mode. Enclose value
in double quotes (“) when setting