Appendix A

Model 6400EH

Teledyne Analytical Instruments

64

SETUP VARIABLE

NUMERIC

UNITS

DEFAULT

VALUE

VALUE RANGE

DESCRIPTION

AUTO, AUTO2

from the RS-232 interface.

PHYS_RANGE1

PPM 500 5–10000

Low

pre-amp

range.

PHYS_RANGE2

PPM

5500

5–10000

High pre-amp range.

CONC_RANGE1

Conc

5000

0.1–50000

D/A concentration range 1.

CONC_RANGE2

Conc

5000

0.1–50000

D/A concentration range 2.

SAMP_FLOW_SET

cc/m 700 0–1200

Sample flow set point for flow
calculation and warning limits.

SAMP_FLOW_SLOPE

1

0.5–1.5

Sample flow slope correction factor
(adjusted flow = measured flow x
slope).

Warnings:
350–1200

VAC_SAMP_RATIO

— 0.53 0.1–2

Maximum vacuum pressure/
sample pressure ratio for valid
sample flow calculation.

SAMP_PRESS_SET

“Hg

29.92

0–100

Sample pressure set point for
pressure compensation and warning
limits.

Warnings

15–35

SAMP PRESS SLOPE

— 1 0.5–1.5

Sample pressure slope correction
factor (adjusted pressure =
measured pressure x slope).

VAC_PRESS_SET

“Hg

6

0-100

Vacuum pressure set point for
pressure compensation and warning
limits.

Warnings:

3-10

BOX SET

°C

30

5-60

Box temperature warning limits.
Set point is not used.

Warnings:

8-50

PMT _SET

ºC

7

0-40

PMT temperature set point and
warning limits.

Warnings:

2–12

RS232_MODE

BitFlag 0 0–65535

RS-232 COM1 mode flags. Add
values to combine flags.

1 = quiet mode

2 = computer mode

4 = enable security

16 = enable Hessen protocol

Must power-cycle instrument for
these options to fully take effect.

32 = enable multi-drop

64 = enable modem

128 = ignore RS-232 line errors

256 = disable XON / XOFF support