Teledyne 6400EH - Sulfur Dioxide Analyzer User Manual

Page 89

Advertising
background image

UV Fluorescence SO2 Analyzer

Appendix A

Teledyne

Analytical

Instruments

77

Appendix A-5: 6400EH iDAS Functions, Revision C.0

Table A-5: 6400EH DAS Trigger Events, Revision C.0

NAME

DESCRIPTION

ATIMER

Automatic timer expired

EXITZR

Exit zero calibration mode

EXITLS

Exit low span calibration mode

EXITHS

Exit high span calibration mode

EXITMP

Exit multi-point calibration mode

SLPCHG

Slope and offset recalculated

EXITDG

Exit diagnostic mode

PMTDTW

PMT detector warning

UVLMPW

UV lamp warning

RCTMPW

Reaction cell temperature warning

PTEMPW

PMT temperature warning

SFLOWW

Sample flow warning

SPRESW

Sample pressure warning

VPRESW

Vacuum pressure warning

BTEMPW

Box temperature warning

HVPSW

High voltage power supply warning

Table A-6: 6400EH iDAS Functions, Revision C.0

NAME

DESCRIPTION

UNITS

PMTDET

PMT detector reading

mV

UVDET

UV lamp intensity reading

mV

LAMPR

UV lamp ratio of calibrated intensity

%

DRKPMT

PMT electrical offset

mV

DARKUV

UV lamp electrical offset

mV

SLOPE1

SO

2

slope for range #1

SLOPE2

SO

2

slope for range #2

OFSET1

SO

2

offset for range #1

mV

OFSET2

SO

2

offset for range #2

mV

Advertising