Troubleshooting Model

6400EH

Teledyne Analytical Instruments

44

If you suspect that one of the two pressure sensors is failing:

1. Cap the sample inlet.
2. After a few seconds, check the

VAC

and

PRES

test functions and

verify that:
 The

VAC

value matches the

PRES

value to within 1 In-Hg-A,

and;

 Both are less than 10 in-Hg-A (i.e. under vacuum).

3. Uncap the sample inlet and unplug the pump.
4. After a few minutes, the value

VAC

and

PRES

should match

within 1 In-Hg-A, and read atmospheric pressure.
 If the two sensors do not match or are significantly different from

ambient atmospheric pressure, call Teledyne Analytical Instruments
customer service.

10.3 Repair Procedures

10.3.1 Repairing the Sample Gas Flow Control Assembly

The Critical Flow Orifice is part of the pressure sensor and flow

control assembly. The jewel orifice is protected by a sintered filter, so it
is unusual for the orifice to need replacing, but it is possible for the
sintered filter and o-rings to need replacing. See the Spare Parts list in
Appendix B for part numbers and kits.

To replace the filter and/or orifice

1. Turn off Power to the analyzer.
2. Locate the pressure sensor / flow control assembly.
3. Disconnect the signal cable and pneumatic fittings.
4. Remove the assembly from the optical bench by removing the 2

screws at each end of the assembly.

5. The inlet end of the assembly is located at the end with the

straight pneumatic fitting. Remove the fitting and the
components as shown in the exploded view.

6. Replace the o-rings (p/n:OR01) and the sintered filter (p/n:

FL01).