Teledyne 6400EH - Sulfur Dioxide Analyzer User Manual

Page 38

Advertising
background image

Operating Instructions

Model 6400EH

Teledyne Analytical Instruments

26

6.2.1 Test Channel Output

When activated, output channel A3 can be used to report one of the

test functions viewable from the SAMPLE mode display. To activate the
A3 channel and select a test function, follow instructions in Section
6.9.10 of the 6400E Manual.

The following table lists test functions that are in addition to or

differ from those listed in Table 6-14 of the 6400E Manual.

Table 6-3: Additional 6400 EH Test Parameters Available for Analog
Output A3

TEST CHANNEL

TEST PARAMETER RANGE

VACUUM PRESSURE

0-40 in-Hg-A

6.2.2 Range Units

The 6400EH only displays concentrations in parts per million (10

6

mols per mol, PPM) or milligrams per cubic meter (mg/m

3

, MGM).

 NOT AVAILABLE: Parts per billion (10

9

mols per mol, PPB)

and micrograms per cubic meter (µg/m

3

, UGM).

To change the concentration units of the 6400EH follow the

instructions found in Section 6.7.7 of the 6400E Manual.

6.2.3 Using the 6400EH with a Hessen Protocol Network

The set up and use of the 6400EH in Hessen protocol networks is

the sane as described in Section 6.12.4 of the M1 00E Manual (P/N
04515) except that there are minor differences in the status flags. The
following table supercedes Table 6-27 of the 6400E Manual.