UV Fluorescence SO2 Analyzer

Appendix A

Teledyne

Analytical

Instruments

65

SETUP VARIABLE

NUMERIC

UNITS

DEFAULT

VALUE

VALUE RANGE

DESCRIPTION

512 = disable hardware FIFOs

1024 = enable RS-485 mode

2048 = even parity, 7 data bits,

1 stop bit

4096 = enable command prompt

BAUD_RATE

— 19200

300,

1200, 2400,
4800, 9600,

19200, 38400,

57600, 115200

RS-232 COM1 baud rate.
Enclose value in double quotes (“)
when setting from the RS-232
interface.

MODEM_INIT

“AT Y0 &D0

&H0 &I0

S0=2 &B0

&N6 &M0 E0

Q1 &W0”

Any character in
the allowed
character set.
Up to 100
characters long.

RS-232 COM 1 modem initialization
string. Sent verbatim plus carriage
return to modem on power up or
manually.

Enclose value in double quotes (“)
when setting from the RS-232
interface.

RS232_MODE2

BitFlag 0 0–65535

RS-232 COM2 mode flags. (Same
settings as RS232_MODE.)

BAUD_RATE2

— 19200

300, 1200,

2400, 4800,

9600, 19200,

38400, 57600,

115200

RS-232 COM2 baud rate. Enclose
value in double quotes (“) when
setting from the RS-232 interface.

MODEM_INIT2

“AT Y0 &D0
&H0 &I0

S0=2 &B0

&N6 &M0 E0

Q1 &W0”

Any character in
the allowed
character set.

Up to 100
characters long.

RS-232 COM2 modem initialization
string. Sent verbatim plus carriage
return to modem on power up or
manually.

Enclose value in double quotes (“)
when setting from the RS-232
interface.

RS232_PASS

Password

940331

0–999999

RS-232 log on password.

MACHINE_ ID

ID

100

0–9999

Unique ID number for instrument.

COMMAND_PROMPT

— “Cmd>

Any character in
the allowed
character set.

Up to 100
characters long.

RS-232 interface command prompt.
Displayed only if enabled with
RS232 MODE variable.

Enclose value in double quotes (“)
when setting from the RS-232
interface.

TEST_CHAN_ID

NONE

NONE,

PMT READING,

UV READING,

VACUUM

Diagnostic analog output ID.

Enclose value in double quotes (“)
when setting from the RS-232
interface.