Teledyne 6400EH - Sulfur Dioxide Analyzer User Manual

Page 60

Advertising
background image

Troubleshooting Model

6400EH

Teledyne Analytical Instruments

48

CAUTION:

ALWAYS WEAR UV-PROTECTIVE, SAFETY
GLASSES WHEN WORKING WITH THE UV LAMP
ASSEMBLY.

1. Set the analyzer display to show the signal I/O function,

UVLAMP_SIGNAL

(see Section 11.1.3).

UVLAMP_SIGNAL

is

function 33.

2. Slightly loosen the large brass thumbscrew located on the shutter

housing (see Figure 10-3) so that the lamp can be moved.

3. While watching the

UVLAMP_SIGNAL

reading, slowly rotate the

lamp or move it back and forth vertically until the

UVLAMP_SIGNAL

reading is at its maximum.

DO NOT grasp the UV lamp by its cap when changing its

position (see Figure 10-3). Always grasp the main body of
the lamp.

4. Compare the

UVLAMP_SIGNAL

reading to the information in

Table 10-3 and follow the instructions there.

Table 10-3: Example of HVPS Power Supply Outputs

UVLAMP_SIGNAL

ACTION TO BE TAKEN

3500mV±200mV.

No Action Required

>

4900mV at any time.

Adjust the UV reference detector potentiometer (see Figure 10-4) until
UVLAMP_SIGNAL

reads approximately 3600mV before continuing to adjust the lamp

position.

>4500mV or < 1000mV

Adjust the UV reference detector potentiometer (see Figure 10-4) until
UVLAMP_SIGNAL

reads as close to 3500mV as possible.

.< 600mV

Replace the lamp.

5. Finger tighten the thumbscrew.

Note: DO NOT over-tighten the thumbscrew.