4 hydrocarbon kicker option (opt 65) – Teledyne 6400EH - Sulfur Dioxide Analyzer User Manual
Page 34

Optional Hardware/Software
Model 6400EH
Teledyne Analytical Instruments
22
Table 5-1: Two-Point Span Valve Operating States
MODE VALVE
CONDITION
SAMPLE
Sample/Cal
Open to SAMPLE inlet
Zero Gas Valve
Closed to ZERO AIR inlet
High Span Valve
Closed to HIGH SPAN inlet
Low Span Valve
Closed to LOW SPAN inlet
ZERO
CAL
Sample/Cal Closed
to SAMPLE inlet
Zero Gas Valve
Open to ZERO AIR inlet
High Span Valve
Closed to HIGH SPAN inlet
Low Span Valve
Closed to LOW SPAN inlet
HIGH
SPAN
CAL
Sample/Cal Closed
to SAMPLE inlet
Zero Gas Valve
Closed to ZERO AIR inlet
High Span Valve
Open to HIGH SPAN inlet
Low Span Valve
Closed to LOW SPAN inlet
LOW
SPAN
CHECK
Sample/Cal Closed
to SAMPLE inlet
Zero Gas Valve
Closed to ZERO AIR inlet
High Span Valve
Closed to HIGH SPAN inlet
Low Span Valve
Open to LOW SPAN inlet
5.4 Hydrocarbon Kicker Option (OPT 65)
This option is specifically designed for those applications where
hydrocarbons are present in the sample gas stream. It includes an
internal, scrubber consisting of a tube of a specialized plastic that
absorbs hydrocarbons very well located within an outer flexible plastic
tube shell.
As gas flows through the inner tube, hydrocarbons are absorbed
into the membrane walls and transported through the membrane wall and
into the hydrocarbon free, purge gas flowing through the outer tube (see
Figure 5-3). This process is driven by the hydrocarbon concentration
gradient between the inner and outer of the tubes.