Optional Hardware/Software

Model 6400EH

Teledyne Analytical Instruments

22

Table 5-1: Two-Point Span Valve Operating States

MODE VALVE

CONDITION

SAMPLE

Sample/Cal

Open to SAMPLE inlet

Zero Gas Valve

Closed to ZERO AIR inlet

High Span Valve

Closed to HIGH SPAN inlet

Low Span Valve

Closed to LOW SPAN inlet

ZERO

CAL

Sample/Cal Closed

to SAMPLE inlet

Zero Gas Valve

Open to ZERO AIR inlet

High Span Valve

Closed to HIGH SPAN inlet

Low Span Valve

Closed to LOW SPAN inlet

HIGH

SPAN

CAL

Sample/Cal Closed

to SAMPLE inlet

Zero Gas Valve

Closed to ZERO AIR inlet

High Span Valve

Open to HIGH SPAN inlet

Low Span Valve

Closed to LOW SPAN inlet

LOW

SPAN

CHECK

Sample/Cal Closed

to SAMPLE inlet

Zero Gas Valve

Closed to ZERO AIR inlet

High Span Valve

Closed to HIGH SPAN inlet

Low Span Valve

Open to LOW SPAN inlet

5.4 Hydrocarbon Kicker Option (OPT 65)

This option is specifically designed for those applications where

hydrocarbons are present in the sample gas stream. It includes an
internal, scrubber consisting of a tube of a specialized plastic that
absorbs hydrocarbons very well located within an outer flexible plastic
tube shell.

As gas flows through the inner tube, hydrocarbons are absorbed

into the membrane walls and transported through the membrane wall and
into the hydrocarbon free, purge gas flowing through the outer tube (see
Figure 5-3). This process is driven by the hydrocarbon concentration
gradient between the inner and outer of the tubes.