Teledyne 6400EH - Sulfur Dioxide Analyzer User Manual
Page 81

UV Fluorescence SO2 Analyzer
Appendix A
Teledyne
Analytical
Instruments
69
Appendix A-3: Warnings and Test Functions,
Revision C.0
Table A-2: 6400EH Warning Messages, Revision C.0
NAME
MESSAGE TEXT
DESCRIPTION
WSYSRES
SYSTEM RESET
Instrument was power-cycled or the CPU was reset.
WDATAINIT
DATA INITIALIZED
Data storage was erased.
WCONFIGINIT
CONFIG INITIALIZED
Configuration storage was reset to factory configuration
or erased.
WPMT
PMT DET WARNING
PMT detector outside of warning limits specified by
DETECTOR_LIMIT
variable.
WUVLAMP
UV LAMP WARNING
UV lamp reading outside of warning limits specified by
DETECTOR_LIMIT
variable.
WSAMPFLOW
SAMPLE FLOW WARN
Sample flow outside of warning limits specified by
SAMP_FLOW_SET
variable.
WSAMPPRESS
SAMPLE PRESS WARN
Sample pressure outside of warning limits specified by
SAMP_PRESS_SET
variable.
WVACPRESS
VACUUM PRESS WARN
Vacuum pressure outside of warning limits specified by
VAC_PRESS_SET
variable.
WBOXTEMP
BOX TEMP WARNING
Chassis temperature outside of warning limits specified
by BOX_SET variable.
WRCELLTEMP
RCELL TEMP WARNING
Reaction cell temperature outside of warning limits
specified by RCELL_SET variable.
WIZSTEMP
IZS TEMP WARNING
IZS temperature outside of warning limits specified by
IZS_SET
variable.
WPMTTEMP
PMT TEMP WARNING
PMT temperature outside of warning limits specified by
PMT_SET
variable.
WDARKCAL
DARK CAL WARNING
Dark offset above limit specified by DARK_LIMIT
variable.
WHVPS
HVPS WARNING
High voltage power supply output outside of warning
limits specified by HVPS_SET variable.
WDYNZERO
CANNOT DYN ZERO
Contact closure zero calibration failed while
DYN_ZERO
was set to ON.
WDYNSPAN
CANNOT DYN SPAN
Contact closure span calibration failed while YN_SPAN
was set to ON.
WREARBOARD
REAR BOARD NOT DET
Rear board was not detected during power up.
WRELAYBOAR
D
RELAY BOARD WARN
Firmware is unable to communicate with the relay
board.