Teledyne 6400EH - Sulfur Dioxide Analyzer User Manual
Page 40

Operating Instructions
Model 6400EH
Teledyne Analytical Instruments
28
6.2.4 Default iDAS Channels
The default Data Channels included in the 6400EH analyzer’s
software include the
CONC, PNUMT
&
CALDAT
channels. The
FAST
&
DETAIL
preset channels are not included.
6.2.5 Remote Operation Using the External Digital I/O
6.2.5.1
S
TATUS
O
UTPUTS
The function and pin assignment5s for the 6400EH digital status
outputs are:.
Table 6-5: Status Output Signals
STATUS
CONNECTOR
PIN NUMBER
1
STATUS
DEFINITION
CONDITION
1
SYSTEM OK
ON if no faults are present.
2
CONC VALID
OFF any time the
HOLD OFF
feature is active, such
as during calibration or when other faults exist
possibly invalidating the current concentration
measurement (example: sample flow rate is outside
of acceptable limits).
ON if concentration measurement is valid.
3
HIGH RANGE
ON if unit is in high range of the
AUTO
Range Mode
4
ZERO CAL
ON whenever the instrument’s ZERO point is being
calibrated.
5
HIGH SPAN CAL
ON whenever the instrument is set for DUAL or
AUTO
reporting range
mode an it’s high range span point is being calibrated
.
6
DIAG MODE
ON whenever the instrument is in
DIAGNO STIC
mode
7
LOW SPAN CAL
ON whenever the instrument is set for DUAL or
AUTO
reporting range mode an it’s lows range span
point is being calibrated .
8
SPARE
D
EMITTER BUS
The emitters of the transistors on pins 1-8 are bussed
together.
SPARE
+
DC POWER
+ 5 VDC, 300 mA source (combined rating with
Control Output, if used).