Teledyne 6400EH - Sulfur Dioxide Analyzer User Manual

Page 40

Advertising
background image

Operating Instructions

Model 6400EH

Teledyne Analytical Instruments

28

6.2.4 Default iDAS Channels

The default Data Channels included in the 6400EH analyzer’s

software include the

CONC, PNUMT

&

CALDAT

channels. The

FAST

&

DETAIL

preset channels are not included.

6.2.5 Remote Operation Using the External Digital I/O

6.2.5.1

S

TATUS

O

UTPUTS

The function and pin assignment5s for the 6400EH digital status

outputs are:.

Table 6-5: Status Output Signals

STATUS

CONNECTOR

PIN NUMBER

1

STATUS

DEFINITION

CONDITION

1

SYSTEM OK

ON if no faults are present.

2

CONC VALID

OFF any time the

HOLD OFF

feature is active, such

as during calibration or when other faults exist
possibly invalidating the current concentration
measurement (example: sample flow rate is outside
of acceptable limits).

ON if concentration measurement is valid.

3

HIGH RANGE

ON if unit is in high range of the

AUTO

Range Mode

4

ZERO CAL

ON whenever the instrument’s ZERO point is being
calibrated.

5

HIGH SPAN CAL

ON whenever the instrument is set for DUAL or
AUTO

reporting range

mode an it’s high range span point is being calibrated
.

6

DIAG MODE

ON whenever the instrument is in

DIAGNO STIC

mode

7

LOW SPAN CAL

ON whenever the instrument is set for DUAL or
AUTO

reporting range mode an it’s lows range span

point is being calibrated .

8

SPARE

D

EMITTER BUS

The emitters of the transistors on pins 1-8 are bussed
together.

SPARE

+

DC POWER

+ 5 VDC, 300 mA source (combined rating with
Control Output, if used).