UV Fluorescence SO2 Analyzer

Appendix A

Teledyne

Analytical

Instruments

75

SIGNAL NAME

BIT OR CHANNEL

NUMBER

DESCRIPTION

SPAN VALVE

7

0 = let span gas in

1 = let zero gas in

LOW_SPAN_VALVE 8

0 = let low span gas in

1 = let sample gas in

ZERO VALVE

9

0 = let zero gas in

1 = let sample gas in

DARKSHUTTER 10

0 = close dark shutter

1 = open

11–15

Spare

Rear board primary MUX analog inputs

PMT_SIGNAL 0

PMT

detector

HVPS_VOLTAGE

1

HV power supply output

PMT_TEMP 2

PMT

temperature

UVLAMP_SIGNAL

3

UV lamp intensity

4

Temperature

MUX

5–6

Spare

SAMPLE_PRESSURE 7

Sample

pressure

TEST_INPUT_8

8

Diagnostic test input

REF_4096_MV

9

4.096V reference from MAX6241

SAMPLE_FLOW 10

Sample flow rate

VACUUM_PRESSURE 10

Vacuum

pressure

TEST_INPUT_11

11

Diagnostic test input

12–13

Spare (thermocouple input?)

14

DAC

MUX

REF_GND 15

Ground

reference

Rear board temperature MUX analog inputs

BOX_TEMP

0

Internal box temperature

RCELL_TEMP

1

Reaction cell temperature

IZS_TEMP 2

IZS

temperature

3

Spare

TEMP_INPUT_4

4

Diagnostic temperature input

TEMP_INPUT_5

5

Diagnostic temperature input