Appendix a model 6400eh, Teledyne analytical instruments – Teledyne 6400EH - Sulfur Dioxide Analyzer User Manual

Page 80

Advertising
background image

Appendix A

Model 6400EH

Teledyne Analytical Instruments

68

SETUP VARIABLE

NUMERIC

UNITS

DEFAULT

VALUE

VALUE RANGE

DESCRIPTION

FACTORY_OPT

BitFlag

0

0–65535

Factory option flags. Add values to
combine flags.

1 = enable dilution factor

2 = zero/span valves installed

4 = IZS installed (implieszero/span

valves installed)

8 = low span valve installed

16 = display units in concentration

field

32 = enable software-controlled

maintenance mode

64 = enable lamp power analog

output

128 = enable switch-controlled

maintenance mode

2048 = enable Internet option

Advertising