Teledyne 6400EH - Sulfur Dioxide Analyzer User Manual

Page 62

Advertising
background image

Troubleshooting Model

6400EH

Teledyne Analytical Instruments

50

Figure 10-5: Pre-Amplifier Board Layout

1. Set the instrument reporting range type to

SNGL

(see Section

6.7.4 of the 6400E Manual)

2. Perform a zero–point calibration using zero air (see Chapter 7 of

the 6400E Manual).

3. Let the instrument stabilize by allowing it to run for one hour.
4. Adjust the UV Lamp. (See Section 10.3.2.1 of this manual)
5. Perform a

LAMP CALIBRATION

procedure (see Section 6.9.7 of

the 6400E Manual).

6. Locate the Preamp board (see Figure 3-4 of this manual).
7. Locate the Following Components On the Preamp board (see

Figure 10-5 of this manual):

HVPS

coarse adjustment switch (Range 0-9, then A-F)

HVPS

fine adjustment switch (Range 0-9, then A-F)

 Gain adjustment potentiometer (Full scale is 10 to 12 turns).

8. Set the

HVPS

coarse adjustment to its minimum setting (0).