REV 1
78
5.0 DIAGNOSTICS, VERIFICATION, AND TROUBLESHOOTING
5.1
Performing a Diagnostics Test
The Diagnostics test mode displays the output voltage at the X leads (V
x
), the voltage sensed by
the H leads (V
h
), and the X voltage excitation current (I
x
). These values can then be verified
using an external volt meter and ampere meter. The Diagnostic Test mode uses the X1-X2 taps.
To perform a diagnostic test:
a. Start from the “START-UP” menu:
Press the
[4]
key (DIAGNOSTIC).
b. The following screen will be displayed:
Select the maximum output voltage on the X terminals by pressing the corresponding
key (
[1]
–
[4]
).
c. The following screen will be displayed:
Press the
[ENTER]
key to continue.
caution! cables will
be energized!
"ENTER" to continue
select voltage range
1.
50v
2.
250v
3.
500v
4.
1500v
1.
RUN TEST 02/16/10
2.
SETUP 10:24:25
3.
TEST PLANS
4.
DIAGNOSTIC